Products

Mouse APOC2(Apolipoprotein C2) ELISA Kit
ELK9918
Size: Price:
48 T $320.00
96 T $458.00

Overview

Product name: Mouse APOC2(Apolipoprotein C2) ELISA Kit
Reactivity: Mouse
Alternative Names: APC2; Apo-C2; ProapoC-II; Apolipoprotein C-II; Proapolipoprotein C-II
Assay Type: Sandwich
Sensitivity: 2.74 ng/mL
Standard: 500 ng/mL
Detection Range: 7.81-500 ng/mL
Sample Type: serum, plasma and other biological fluids
Assay Length: 3.5h
Research Area: Metabolic pathway;Cardiovascular biology;Hepatology;Nutrition metabolism;
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Mouse APOC2. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Mouse APOC2. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Mouse APOC2, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Mouse APOC2 in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (ng/mL) OD Corrected OD
500.00 2.204 2.108
250.00 1.527 1.431
125.00 1.115 1.019
62.50 0.661 0.565
31.25 0.532 0.436
15.63 0.385 0.289
7.82 0.208 0.112
0.00 0.096 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant APOC2 and the recovery rates were calculated by comparing the measured value to the expected amount of APOC2 in samples.
Matrix Recovery range Average
serum(n=5) 82-95% 88%
EDTA plasma(n=5) 90-105% 97%
Heparin plasma(n=5) 87-99% 93%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of APOC2 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 90-99% 95-106% 87-101% 82-97%
EDTA plasma(n=5) 91-99% 88-97% 88-104% 97-108%
Heparin plasma(n=5) 95-103% 87-98% 85-92% 79-96%