| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 100.00 | 1.881 | 1.806 |
| 50.00 | 1.588 | 1.513 |
| 25.00 | 1.178 | 1.103 |
| 12.50 | 0.735 | 0.660 |
| 6.25 | 0.533 | 0.458 |
| 3.13 | 0.369 | 0.294 |
| 1.56 | 0.227 | 0.152 |
| 0.00 | 0.075 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 91-104% | 98% |
| EDTA plasma(n=5) | 86-128% | 107% |
| Heparin plasma(n=5) | 87-101% | 94% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 83-107% | 87-104% | 86-117% | 92-120% |
| EDTA plasma(n=5) | 81-114% | 95-106% | 82-96% | 82-103% |
| Heparin plasma(n=5) | 88-96% | 88-105% | 91-110% | 85-102% |