| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.172 | 2.076 |
| 500.00 | 1.543 | 1.447 |
| 250.00 | 1.277 | 1.181 |
| 125.00 | 0.952 | 0.856 |
| 62.50 | 0.496 | 0.400 |
| 31.25 | 0.305 | 0.209 |
| 15.63 | 0.194 | 0.098 |
| 0.00 | 0.096 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-107% | 98% |
| EDTA plasma(n=5) | 92-107% | 99% |
| Heparin plasma(n=5) | 89-103% | 96% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 91-102% | 79-91% | 89-103% | 82-95% |
| EDTA plasma(n=5) | 90-99% | 87-103% | 81-96% | 86-93% |
| Heparin plasma(n=5) | 86-97% | 83-96% | 91-103% | 89-96% |