Size: | Price: |
---|---|
48 T | $265.00 |
96 T | $376.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
10.00 | 1.842 | 1.755 |
5.00 | 1.572 | 1.485 |
2.50 | 1.227 | 1.140 |
1.25 | 0.852 | 0.765 |
0.63 | 0.508 | 0.421 |
0.32 | 0.357 | 0.270 |
0.16 | 0.215 | 0.128 |
0.00 | 0.087 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 82-94% | 88% |
EDTA plasma(n=5) | 78-91% | 84% |
Heparin plasma(n=5) | 83-97% | 90% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 87-98% | 86-102% | 87-99% | 99-105% |
EDTA plasma(n=5) | 88-97% | 93-103% | 81-98% | 78-90% |
Heparin plasma(n=5) | 91-99% | 97-103% | 81-96% | 87-105% |