| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 500.00 | 2.094 | 2.000 | 
| 250.00 | 1.543 | 1.449 | 
| 125.00 | 1.240 | 1.146 | 
| 62.50 | 0.892 | 0.798 | 
| 31.25 | 0.582 | 0.488 | 
| 15.63 | 0.328 | 0.234 | 
| 7.81 | 0.231 | 0.137 | 
| 0.00 | 0.094 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 81-101% | 91% | 
| EDTA plasma(n=5) | 89-103% | 96% | 
| Heparin plasma(n=5) | 83-113% | 98% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 84-100% | 94-112% | 83-115% | 87-97% | 
| EDTA plasma(n=5) | 90-114% | 84-115% | 92-108% | 94-105% | 
| Heparin plasma(n=5) | 86-95% | 86-116% | 88-112% | 91-99% | 