| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 8000.00 | 2.096 | 2.028 | 
| 4000.00 | 1.648 | 1.580 | 
| 2000.00 | 1.096 | 1.028 | 
| 1000.00 | 0.852 | 0.784 | 
| 500.00 | 0.491 | 0.423 | 
| 250.00 | 0.369 | 0.301 | 
| 125.00 | 0.251 | 0.183 | 
| 0.00 | 0.068 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 91-107% | 99% | 
| EDTA plasma(n=5) | 90-112% | 101% | 
| Heparin plasma(n=5) | 91-106% | 99% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 81-90% | 92-113% | 80-87% | 81-102% | 
| EDTA plasma(n=5) | 90-108% | 84-109% | 91-98% | 91-109% | 
| Heparin plasma(n=5) | 87-96% | 82-113% | 91-119% | 87-117% | 