Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
100.00 | 2.154 | 2.068 |
50.00 | 1.686 | 1.600 |
25.00 | 1.211 | 1.125 |
12.50 | 0.913 | 0.827 |
6.25 | 0.585 | 0.499 |
3.13 | 0.355 | 0.269 |
1.57 | 0.189 | 0.103 |
0.00 | 0.086 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 80-95% | 87% |
EDTA plasma(n=5) | 85-97% | 91% |
Heparin plasma(n=5) | 92-107% | 99% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 87-96% | 85-97% | 85-94% | 84-98% |
EDTA plasma(n=5) | 82-97% | 88-105% | 79-93% | 87-98% |
Heparin plasma(n=5) | 89-96% | 79-98% | 81-92% | 87-101% |