| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.270 | 2.182 |
| 1000.00 | 1.724 | 1.636 |
| 500.00 | 1.177 | 1.089 |
| 250.00 | 0.865 | 0.777 |
| 125.00 | 0.548 | 0.460 |
| 62.50 | 0.371 | 0.283 |
| 31.25 | 0.216 | 0.128 |
| 0.00 | 0.088 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 91-111% | 101% |
| EDTA plasma(n=5) | 88-110% | 99% |
| Heparin plasma(n=5) | 94-104% | 99% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-117% | 87-94% | 82-104% | 81-102% |
| EDTA plasma(n=5) | 87-110% | 89-108% | 81-119% | 90-116% |
| Heparin plasma(n=5) | 92-115% | 90-100% | 81-98% | 90-118% |