Products

Horse BMP2(Bone Morphogenetic Protein 2) ELISA Kit
ELK8969
Size: Price:
48 T $412.00
96 T $588.00

Overview

Product name: Horse BMP2(Bone Morphogenetic Protein 2) ELISA Kit
Reactivity: Horse
Alternative Names: BMP2A; BMP-2A; Hemochromatosis Modifier
Assay Type: Sandwich
Sensitivity: 13.1 pg/mL
Standard: 2000 pg/mL
Detection Range: 31.25-2000 pg/mL
Sample Type: Serum, plasma, tissue homogenates and other biological fluids.
Assay Length: 3.5h
Research Area: Developmental science;Bone metabolism;
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Horse BMP2. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Horse BMP2. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Horse BMP2, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Horse BMP2 in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (pg/mL) OD Corrected OD
2000.00 2.075 1.992
1000.00 1.711 1.628
500.00 1.116 1.033
250.00 0.847 0.764
125.00 0.509 0.426
62.50 0.357 0.274
31.25 0.196 0.113
0.00 0.083 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant BMP2 and the recovery rates were calculated by comparing the measured value to the expected amount of BMP2 in samples.
Matrix Recovery range Average
serum(n=5) 87-95% 91%
EDTA plasma(n=5) 93-107% 100%
Heparin plasma(n=5) 87-99% 93%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of BMP2 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 81-92% 92-103% 87-98% 90-101%
EDTA plasma(n=5) 89-97% 85-94% 79-96% 85-99%
Heparin plasma(n=5) 87-102% 96-105% 85-93% 89-97%