| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 300.00 | 2.427 | 2.355 |
| 150.00 | 1.599 | 1.527 |
| 75.00 | 1.114 | 1.042 |
| 37.50 | 0.898 | 0.826 |
| 18.75 | 0.570 | 0.498 |
| 9.38 | 0.321 | 0.249 |
| 4.69 | 0.194 | 0.122 |
| 0.00 | 0.072 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 94-117% | 106% |
| EDTA plasma(n=5) | 80-96% | 88% |
| Heparin plasma(n=5) | 88-114% | 101% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 92-112% | 80-106% | 91-96% | 92-100% |
| EDTA plasma(n=5) | 86-96% | 91-114% | 86-94% | 94-102% |
| Heparin plasma(n=5) | 81-100% | 87-109% | 82-90% | 93-120% |