| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 20.00 | 2.168 | 2.095 |
| 10.00 | 1.782 | 1.709 |
| 5.00 | 1.052 | 0.979 |
| 2.50 | 0.797 | 0.724 |
| 1.25 | 0.567 | 0.494 |
| 0.63 | 0.315 | 0.242 |
| 0.32 | 0.142 | 0.069 |
| 0.00 | 0.073 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 81-95% | 88% |
| EDTA plasma(n=5) | 83-95% | 89% |
| Heparin plasma(n=5) | 87-97% | 90% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 92-103% | 93-105% | 88-101% | 97-106% |
| EDTA plasma(n=5) | 85-92% | 90-96% | 87-1028% | 85-97% |
| Heparin plasma(n=5) | 96-105% | 95-103% | 95-103% | 87-98% |