| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 1.947 | 1.866 |
| 5.00 | 1.553 | 1.472 |
| 2.50 | 1.204 | 1.123 |
| 1.25 | 0.828 | 0.747 |
| 0.63 | 0.538 | 0.457 |
| 0.31 | 0.357 | 0.276 |
| 0.16 | 0.228 | 0.147 |
| 0.00 | 0.081 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 92-109% | 101% |
| EDTA plasma(n=5) | 91-128% | 110% |
| Heparin plasma(n=5) | 94-112% | 103% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 90-103% | 93-111% | 94-104% | 93-109% |
| EDTA plasma(n=5) | 83-110% | 89-114% | 87-103% | 94-117% |
| Heparin plasma(n=5) | 88-97% | 92-118% | 86-101% | 86-101% |