48 T | $412.00 |
96 T | $588.00 |
Concentration (pg/mL) | OD | Corrected OD |
---|---|---|
1000.00 | 2.201 | 2.118 |
500.00 | 1.407 | 1.324 |
250.00 | 1.043 | 0.960 |
125.00 | 0.782 | 0.699 |
62.50 | 0.595 | 0.512 |
31.25 | 0.365 | 0.282 |
15.63 | 0.179 | 0.096 |
0.00 | 0.083 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 94-107% | 100% |
EDTA plasma(n=5) | 94-105% | 98% |
Heparin plasma(n=5) | 82-95% | 88% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 93-102% | 81-96% | 93-106% | 82-95% |
EDTA plasma(n=5) | 89-97% | 95-105% | 82-95% | 81-97% |
Heparin plasma(n=5) | 82-96% | 95-102% | 89-101% | 87-98% |