| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 3000.00 | 1.942 | 1.855 |
| 1500.00 | 1.617 | 1.530 |
| 750.00 | 1.176 | 1.089 |
| 375.00 | 0.893 | 0.806 |
| 187.50 | 0.527 | 0.440 |
| 93.75 | 0.355 | 0.268 |
| 46.88 | 0.187 | 0.100 |
| 0.00 | 0.087 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-116% | 102% |
| EDTA plasma(n=5) | 81-112% | 97% |
| Heparin plasma(n=5) | 92-109% | 101% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-104% | 84-119% | 88-115% | 90-115% |
| EDTA plasma(n=5) | 88-112% | 88-117% | 81-107% | 92-104% |
| Heparin plasma(n=5) | 90-116% | 86-120% | 88-119% | 91-103% |