Size: | Price: |
---|---|
48 T | $412.00 |
96 T | $588.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
100.00 | 1.916 | 1.840 |
50.00 | 1.536 | 1.460 |
25.00 | 1.183 | 1.107 |
12.50 | 0.782 | 0.706 |
6.25 | 0.545 | 0.469 |
3.13 | 0.357 | 0.281 |
1.57 | 0.236 | 0.160 |
0.00 | 0.076 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 85-99% | 92% |
EDTA plasma(n=5) | 93-107% | 100% |
Heparin plasma(n=5) | 87-99% | 93% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 87-98% | 87-100% | 85-94% | 83-96% |
EDTA plasma(n=5) | 90-99% | 85-96% | 87-98% | 93-102% |
Heparin plasma(n=5) | 85-93% | 87-96% | 79-91% | 95-103% |