| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.079 | 1.993 |
| 5.00 | 1.698 | 1.612 |
| 2.50 | 1.140 | 1.054 |
| 1.25 | 0.886 | 0.800 |
| 0.63 | 0.607 | 0.521 |
| 0.31 | 0.361 | 0.275 |
| 0.16 | 0.214 | 0.128 |
| 0.00 | 0.086 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 88-105% | 97% |
| EDTA plasma(n=5) | 94-101% | 98% |
| Heparin plasma(n=5) | 80-95% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-111% | 94-106% | 94-116% | 89-103% |
| EDTA plasma(n=5) | 94-109% | 81-92% | 83-118% | 90-105% |
| Heparin plasma(n=5) | 92-100% | 91-99% | 88-114% | 85-113% |