| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.146 | 2.051 |
| 5.00 | 1.589 | 1.494 |
| 2.50 | 1.136 | 1.041 |
| 1.25 | 0.872 | 0.777 |
| 0.63 | 0.468 | 0.373 |
| 0.31 | 0.320 | 0.225 |
| 0.16 | 0.213 | 0.118 |
| 0.00 | 0.095 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 82-96% | 89% |
| EDTA plasma(n=5) | 85-121% | 103% |
| Heparin plasma(n=5) | 85-111% | 98% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 94-114% | 80-108% | 80-88% | 88-98% |
| EDTA plasma(n=5) | 94-120% | 93-119% | 86-106% | 84-93% |
| Heparin plasma(n=5) | 88-100% | 86-101% | 86-96% | 90-110% |