| 48 T | $265.00 | 
| 96 T | $376.00 | 
| 96 T*5 | $1598.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 900.00 | 0.235 | |
| 450.00 | 0.432 | |
| 225.00 | 0.655 | |
| 112.50 | 0.895 | |
| 56.25 | 1.189 | |
| 28.13 | 1.579 | |
| 14.07 | 1.733 | |
| 0.00 | 2.028 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 80-95% | 87% | 
| EDTA plasma(n=5) | 78-96% | 87% | 
| Heparin plasma(n=5) | 78-92% | 87% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 79-93% | 89-97% | 92-105% | 83-92% | 
| EDTA plasma(n=5) | 87-96% | 85-98% | 89-102% | 85-98% | 
| Heparin plasma(n=5) | 79-93% | 82-96% | 92-105% | 83-92% |