| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 500.00 | 2.379 | 2.290 | 
| 250.00 | 1.565 | 1.476 | 
| 125.00 | 1.196 | 1.107 | 
| 62.50 | 0.797 | 0.708 | 
| 31.25 | 0.508 | 0.419 | 
| 15.63 | 0.336 | 0.247 | 
| 7.82 | 0.198 | 0.109 | 
| 0.00 | 0.089 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 85-103% | 94% | 
| EDTA plasma(n=5) | 95-105% | 100% | 
| Heparin plasma(n=5) | 93-101% | 97% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 88-97% | 78-96% | 80-99% | 89-97% | 
| EDTA plasma(n=5) | 86-93% | 91-105% | 87-98% | 89-94% | 
| Heparin plasma(n=5) | 93-102% | 93-106% | 87-101% | 91-104% |