| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 200.00 | 0.211 | |
| 100.00 | 0.389 | |
| 50.00 | 0.621 | |
| 25.00 | 0.861 | |
| 12.50 | 1.152 | |
| 6.25 | 1.565 | |
| 3.13 | 1.789 | |
| 0.00 | 2.123 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-99% | 95% |
| EDTA plasma(n=5) | 85-97% | 91% |
| Heparin plasma(n=5) | 85-97% | 92% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 91-99% | 83-97% | 98-105% | 96-101% |
| EDTA plasma(n=5) | 85-92% | 79-96% | 89-101% | 85-97% |
| Heparin plasma(n=5) | 92-105% | 86-98% | 80-93% | 89-97% |