| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 1.853 | 1.750 |
| 500.00 | 1.578 | 1.475 |
| 250.00 | 1.115 | 1.012 |
| 125.00 | 0.830 | 0.727 |
| 62.50 | 0.575 | 0.472 |
| 31.25 | 0.380 | 0.277 |
| 15.63 | 0.305 | 0.202 |
| 0.00 | 0.103 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 92-103% | 98% |
| EDTA plasma(n=5) | 83-101% | 92% |
| Heparin plasma(n=5) | 83-113% | 98% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 94-99% | 88-95% | 93-115% | 81-119% |
| EDTA plasma(n=5) | 91-97% | 91-100% | 84-91% | 93-117% |
| Heparin plasma(n=5) | 81-100% | 93-115% | 84-93% | 88-104% |