| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.172 | 2.085 |
| 500.00 | 1.604 | 1.517 |
| 250.00 | 1.151 | 1.064 |
| 125.00 | 0.916 | 0.829 |
| 62.50 | 0.570 | 0.483 |
| 31.25 | 0.318 | 0.231 |
| 15.63 | 0.190 | 0.103 |
| 0.00 | 0.087 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 92-99% | 96% |
| EDTA plasma(n=5) | 92-119% | 106% |
| Heparin plasma(n=5) | 89-99% | 94% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 95-114% | 90-119% | 87-103% | 81-116% |
| EDTA plasma(n=5) | 92-117% | 94-112% | 93-106% | 95-109% |
| Heparin plasma(n=5) | 90-101% | 90-107% | 88-117% | 92-98% |