| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 4000.00 | 1.978 | 1.883 |
| 2000.00 | 1.636 | 1.541 |
| 1000.00 | 1.135 | 1.040 |
| 500.00 | 0.775 | 0.680 |
| 250.00 | 0.487 | 0.392 |
| 125.00 | 0.322 | 0.227 |
| 62.50 | 0.238 | 0.143 |
| 0.00 | 0.095 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 82-90% | 86% |
| EDTA plasma(n=5) | 84-92% | 87% |
| Heparin plasma(n=5) | 95-104% | 98% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-96% | 90-101% | 93-102% | 85-94% |
| EDTA plasma(n=5) | 85-94% | 79-92% | 87-101% | 82-90% |
| Heparin plasma(n=5) | 97-105% | 84-96% | 87-98% | 95-102% |