| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.071 | 1.983 |
| 5.00 | 1.716 | 1.628 |
| 2.50 | 0.959 | 0.871 |
| 1.25 | 0.782 | 0.694 |
| 0.63 | 0.548 | 0.460 |
| 0.31 | 0.321 | 0.233 |
| 0.16 | 0.240 | 0.152 |
| 0.00 | 0.088 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 91-105% | 98% |
| EDTA plasma(n=5) | 91-115% | 103% |
| Heparin plasma(n=5) | 84-96% | 90% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 93-109% | 83-89% | 84-101% | 86-94% |
| EDTA plasma(n=5) | 90-106% | 85-116% | 86-107% | 91-117% |
| Heparin plasma(n=5) | 92-111% | 90-104% | 91-108% | 85-114% |