| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 1.892 | 1.799 |
| 5.00 | 1.680 | 1.587 |
| 2.50 | 1.196 | 1.103 |
| 1.25 | 0.792 | 0.699 |
| 0.63 | 0.526 | 0.433 |
| 0.31 | 0.334 | 0.241 |
| 0.16 | 0.220 | 0.127 |
| 0.00 | 0.093 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-93% | 87% |
| EDTA plasma(n=5) | 82-123% | 103% |
| Heparin plasma(n=5) | 81-101% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-102% | 89-108% | 88-119% | 94-105% |
| EDTA plasma(n=5) | 86-91% | 95-114% | 85-105% | 82-119% |
| Heparin plasma(n=5) | 89-117% | 86-100% | 89-96% | 82-91% |