| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.120 | 2.012 |
| 5.00 | 1.573 | 1.465 |
| 2.50 | 1.165 | 1.057 |
| 1.25 | 0.791 | 0.683 |
| 0.63 | 0.545 | 0.437 |
| 0.31 | 0.317 | 0.209 |
| 0.16 | 0.188 | 0.080 |
| 0.00 | 0.108 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-109% | 100% |
| EDTA plasma(n=5) | 87-100% | 94% |
| Heparin plasma(n=5) | 85-91% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 84-99% | 89-98% | 94-107% | 82-116% |
| EDTA plasma(n=5) | 82-113% | 83-115% | 91-118% | 80-99% |
| Heparin plasma(n=5) | 85-110% | 83-90% | 92-118% | 86-115% |