| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.126 | 2.047 |
| 5.00 | 1.588 | 1.509 |
| 2.50 | 1.176 | 1.097 |
| 1.25 | 0.827 | 0.748 |
| 0.63 | 0.527 | 0.448 |
| 0.31 | 0.244 | 0.165 |
| 0.16 | 0.188 | 0.109 |
| 0.00 | 0.079 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 94-111% | 103% |
| EDTA plasma(n=5) | 92-119% | 106% |
| Heparin plasma(n=5) | 95-115% | 105% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-120% | 93-114% | 89-115% | 92-114% |
| EDTA plasma(n=5) | 91-98% | 89-116% | 94-112% | 93-107% |
| Heparin plasma(n=5) | 94-108% | 87-94% | 90-108% | 83-112% |