| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 2.167 | 2.075 |
| 2500.00 | 1.573 | 1.481 |
| 1250.00 | 1.041 | 0.949 |
| 625.00 | 0.923 | 0.831 |
| 312.50 | 0.632 | 0.540 |
| 156.25 | 0.320 | 0.228 |
| 78.13 | 0.153 | 0.061 |
| 0.00 | 0.092 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-109% | 100% |
| EDTA plasma(n=5) | 93-123% | 108% |
| Heparin plasma(n=5) | 93-101% | 97% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-105% | 84-90% | 93-112% | 85-108% |
| EDTA plasma(n=5) | 90-119% | 85-98% | 87-111% | 84-113% |
| Heparin plasma(n=5) | 94-107% | 88-102% | 87-110% | 90-118% |