| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 1.885 | 1.799 |
| 500.00 | 1.730 | 1.644 |
| 250.00 | 1.075 | 0.989 |
| 125.00 | 0.753 | 0.667 |
| 62.50 | 0.547 | 0.461 |
| 31.25 | 0.329 | 0.243 |
| 15.63 | 0.246 | 0.160 |
| 0.00 | 0.086 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 81-96% | 89% |
| EDTA plasma(n=5) | 93-129% | 111% |
| Heparin plasma(n=5) | 91-111% | 101% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-107% | 89-116% | 89-115% | 82-118% |
| EDTA plasma(n=5) | 89-101% | 84-94% | 84-91% | 92-106% |
| Heparin plasma(n=5) | 89-114% | 85-106% | 86-92% | 94-119% |