| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 200.00 | 1.972 | 1.899 |
| 100.00 | 1.545 | 1.472 |
| 50.00 | 1.136 | 1.063 |
| 25.00 | 0.975 | 0.902 |
| 12.50 | 0.509 | 0.436 |
| 6.25 | 0.377 | 0.304 |
| 3.13 | 0.265 | 0.192 |
| 0.00 | 0.073 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 81-95% | 88% |
| EDTA plasma(n=5) | 95-103% | 99% |
| Heparin plasma(n=5) | 92-102% | 97% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 90-99% | 83-97% | 81-103% | 89-101% |
| EDTA plasma(n=5) | 82-94% | 85-96% | 78-90% | 91-99% |
| Heparin plasma(n=5) | 85-94% | 87-101% | 86-97% | 88-95% |