| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.240 | 2.148 |
| 5.00 | 1.502 | 1.410 |
| 2.50 | 1.201 | 1.109 |
| 1.25 | 0.959 | 0.867 |
| 0.63 | 0.587 | 0.495 |
| 0.31 | 0.296 | 0.204 |
| 0.16 | 0.170 | 0.078 |
| 0.00 | 0.092 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 84-98% | 91% |
| EDTA plasma(n=5) | 92-106% | 99% |
| Heparin plasma(n=5) | 86-104% | 95% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-115% | 81-100% | 86-106% | 95-115% |
| EDTA plasma(n=5) | 95-111% | 86-91% | 82-90% | 80-87% |
| Heparin plasma(n=5) | 93-112% | 84-103% | 84-106% | 90-98% |