| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 4000.00 | 1.850 | 1.747 |
| 2000.00 | 1.526 | 1.423 |
| 1000.00 | 1.245 | 1.142 |
| 500.00 | 0.866 | 0.763 |
| 250.00 | 0.520 | 0.417 |
| 125.00 | 0.326 | 0.223 |
| 62.50 | 0.219 | 0.116 |
| 0.00 | 0.103 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-89% | 85% |
| EDTA plasma(n=5) | 82-88% | 85% |
| Heparin plasma(n=5) | 91-116% | 104% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 84-119% | 83-115% | 90-110% | 82-118% |
| EDTA plasma(n=5) | 88-111% | 82-113% | 87-105% | 84-118% |
| Heparin plasma(n=5) | 86-117% | 85-94% | 84-96% | 92-119% |