Products

Rabbit cTnT/TNNT2(Troponin T Type 2, Cardiac) ELISA Kit
ELK6945
Size: Price:
48 T $412.00
96 T $588.00

Overview

Product name: Rabbit cTnT/TNNT2(Troponin T Type 2, Cardiac) ELISA Kit
Reactivity: Rabbit
Alternative Names: CMH2; cTnT; TnTc; Cardiomyopathy,Hypertrophic 2; Cardiac muscle troponin T; TNNT2; Troponin T Type 2, Cardiac
Assay Type: Sandwich
Sensitivity: 6 pg/mL
Standard: 1000 pg/mL
Detection Range: 15.63-1000 pg/mL
Sample Type: Serum, plasma and other biological fluids.
Assay Length: 3.5h
Research Area: Cardiovascular biology;
Uniprot ID: P09741
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Rabbit cTnT/TNNT2. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Rabbit cTnT/TNNT2. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Rabbit cTnT/TNNT2, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Rabbit cTnT/TNNT2 in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (pg/mL) OD Corrected OD
1000.00 2.135 2.042
500.00 1.562 1.469
250.00 0.913 0.820
125.00 0.682 0.589
62.50 0.457 0.364
31.25 0.239 0.146
15.63 0.228 0.135
0.00 0.093 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant cTnT/TNNT2 and the recovery rates were calculated by comparing the measured value to the expected amount of cTnT/TNNT2 in samples.
Matrix Recovery range Average
serum(n=5) 87-93% 90%
EDTA plasma(n=5) 84-106% 100%
Heparin plasma(n=5) 87-99% 93%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of cTnT/TNNT2 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 85-105% 87-98% 91-102% 86-102%
EDTA plasma(n=5) 79-97% 96-104% 85-93% 89-97%
Heparin plasma(n=5) 85-94% 87-103% 97-105% 79-90%