| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.059 | 1.971 |
| 500.00 | 1.533 | 1.445 |
| 250.00 | 1.211 | 1.123 |
| 125.00 | 0.751 | 0.663 |
| 62.50 | 0.456 | 0.368 |
| 31.25 | 0.369 | 0.281 |
| 15.63 | 0.207 | 0.119 |
| 0.00 | 0.088 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 89-117% | 103% |
| EDTA plasma(n=5) | 95-115% | 105% |
| Heparin plasma(n=5) | 86-111% | 99% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-112% | 82-92% | 95-110% | 87-98% |
| EDTA plasma(n=5) | 94-119% | 80-117% | 95-107% | 93-98% |
| Heparin plasma(n=5) | 89-100% | 82-95% | 90-108% | 88-119% |