| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 2.127 | 2.034 |
| 2500.00 | 1.740 | 1.647 |
| 1250.00 | 1.263 | 1.170 |
| 625.00 | 0.834 | 0.741 |
| 312.50 | 0.574 | 0.481 |
| 156.25 | 0.375 | 0.282 |
| 78.13 | 0.188 | 0.095 |
| 0.00 | 0.093 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-110% | 100% |
| EDTA plasma(n=5) | 87-130% | 109% |
| Heparin plasma(n=5) | 95-118% | 107% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-96% | 87-115% | 82-97% | 92-119% |
| EDTA plasma(n=5) | 86-92% | 89-103% | 81-115% | 81-92% |
| Heparin plasma(n=5) | 86-100% | 88-116% | 86-117% | 90-118% |