Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (pg/mL) | OD | Corrected OD |
---|---|---|
1000.00 | 1.990 | 1.904 |
500.00 | 1.653 | 1.567 |
250.00 | 1.152 | 1.066 |
125.00 | 0.892 | 0.806 |
62.50 | 0.547 | 0.461 |
31.25 | 0.373 | 0.287 |
15.63 | 0.223 | 0.137 |
0.00 | 0.086 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 86-99% | 92% |
EDTA plasma(n=5) | 80-95% | 87% |
Heparin plasma(n=5) | 90-105% | 97% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 87-98% | 92-105% | 83-92% | 89-98% |
EDTA plasma(n=5) | 85-102% | 81-96% | 87-105% | 83-92% |
Heparin plasma(n=5) | 88-97% | 86-92% | 79-103% | 89-97% |