| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.182 | 2.094 |
| 5.00 | 1.621 | 1.533 |
| 2.50 | 1.097 | 1.009 |
| 1.25 | 0.711 | 0.623 |
| 0.63 | 0.576 | 0.488 |
| 0.31 | 0.388 | 0.300 |
| 0.16 | 0.165 | 0.077 |
| 0.00 | 0.088 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 81-88% | 85% |
| EDTA plasma(n=5) | 82-118% | 100% |
| Heparin plasma(n=5) | 93-100% | 97% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 95-117% | 92-101% | 87-98% | 83-95% |
| EDTA plasma(n=5) | 93-106% | 95-104% | 81-115% | 93-111% |
| Heparin plasma(n=5) | 86-93% | 86-101% | 92-119% | 91-111% |