Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (pg/mL) | OD | Corrected OD |
---|---|---|
1000.00 | 2.117 | 2.049 |
500.00 | 1.593 | 1.525 |
250.00 | 1.088 | 1.020 |
125.00 | 0.682 | 0.614 |
62.50 | 0.539 | 0.471 |
31.25 | 0.335 | 0.267 |
15.63 | 0.226 | 0.158 |
0.00 | 0.068 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 80-95% | 87% |
EDTA plasma(n=5) | 92-105% | 98% |
Heparin plasma(n=5) | 84-96% | 90% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 85-92% | 80-96% | 87-98% | 95-102% |
EDTA plasma(n=5) | 87-95% | 78-91% | 97-103% | 95-101% |
Heparin plasma(n=5) | 89-103% | 79-92% | 88-97% | 85-97% |