| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 20.00 | 2.072 | 1.978 |
| 10.00 | 1.740 | 1.646 |
| 5.00 | 1.246 | 1.152 |
| 2.50 | 0.920 | 0.826 |
| 1.25 | 0.531 | 0.437 |
| 0.63 | 0.294 | 0.200 |
| 0.31 | 0.217 | 0.123 |
| 0.00 | 0.094 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 86-92% | 89% |
| EDTA plasma(n=5) | 86-111% | 99% |
| Heparin plasma(n=5) | 87-117% | 102% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 90-95% | 94-106% | 90-113% | 80-95% |
| EDTA plasma(n=5) | 87-97% | 80-120% | 93-118% | 87-105% |
| Heparin plasma(n=5) | 83-118% | 90-107% | 88-107% | 90-105% |