| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 8000.00 | 2.089 | 2.006 |
| 4000.00 | 1.713 | 1.630 |
| 2000.00 | 1.209 | 1.126 |
| 1000.00 | 0.934 | 0.851 |
| 500.00 | 0.520 | 0.437 |
| 250.00 | 0.409 | 0.326 |
| 125.00 | 0.190 | 0.107 |
| 0.00 | 0.083 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 89-95% | 92% |
| EDTA plasma(n=5) | 91-126% | 109% |
| Heparin plasma(n=5) | 85-94% | 90% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 92-119% | 93-103% | 84-106% | 89-101% |
| EDTA plasma(n=5) | 90-116% | 87-108% | 95-103% | 88-95% |
| Heparin plasma(n=5) | 93-105% | 81-90% | 94-100% | 94-103% |