Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
10.00 | 2.201 | 2.113 |
5.00 | 1.593 | 1.505 |
2.50 | 1.144 | 1.056 |
1.25 | 0.892 | 0.804 |
0.63 | 0.492 | 0.404 |
0.32 | 0.332 | 0.244 |
0.16 | 0.168 | 0.080 |
0.00 | 0.088 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 87-99% | 93% |
EDTA plasma(n=5) | 93-107% | 100% |
Heparin plasma(n=5) | 87-99% | 93% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 91-99% | 88-97% | 88-104% | 97-108% |
EDTA plasma(n=5) | 86-94% | 93-102% | 87-96% | 92-105% |
Heparin plasma(n=5) | 80-101% | 87-96% | 96-105% | 95-103% |