| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 800.00 | 2.167 | 2.086 |
| 400.00 | 1.700 | 1.619 |
| 200.00 | 1.209 | 1.128 |
| 100.00 | 0.833 | 0.752 |
| 50.00 | 0.605 | 0.524 |
| 25.00 | 0.319 | 0.238 |
| 12.50 | 0.258 | 0.177 |
| 0.00 | 0.081 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-93% | 88% |
| EDTA plasma(n=5) | 87-110% | 99% |
| Heparin plasma(n=5) | 81-97% | 89% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-116% | 93-113% | 84-89% | 86-95% |
| EDTA plasma(n=5) | 86-99% | 94-117% | 86-103% | 87-106% |
| Heparin plasma(n=5) | 87-112% | 92-108% | 88-116% | 88-105% |