| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 20.00 | 2.219 | 2.121 |
| 10.00 | 1.474 | 1.376 |
| 5.00 | 1.126 | 1.028 |
| 2.50 | 0.999 | 0.901 |
| 1.25 | 0.465 | 0.367 |
| 0.63 | 0.319 | 0.221 |
| 0.31 | 0.212 | 0.114 |
| 0.00 | 0.098 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-108% | 99% |
| EDTA plasma(n=5) | 83-108% | 96% |
| Heparin plasma(n=5) | 82-116% | 99% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 81-94% | 82-112% | 85-91% | 95-120% |
| EDTA plasma(n=5) | 88-108% | 88-105% | 94-106% | 85-96% |
| Heparin plasma(n=5) | 86-100% | 83-113% | 89-100% | 90-100% |