| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.026 | 1.934 |
| 5.00 | 1.653 | 1.561 |
| 2.50 | 1.126 | 1.034 |
| 1.25 | 0.885 | 0.793 |
| 0.63 | 0.554 | 0.462 |
| 0.31 | 0.389 | 0.297 |
| 0.16 | 0.216 | 0.124 |
| 0.00 | 0.092 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-100% | 92% |
| EDTA plasma(n=5) | 88-129% | 109% |
| Heparin plasma(n=5) | 91-110% | 101% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 81-98% | 91-110% | 93-108% | 80-118% |
| EDTA plasma(n=5) | 88-93% | 92-117% | 91-103% | 84-118% |
| Heparin plasma(n=5) | 89-103% | 86-104% | 88-97% | 91-117% |