| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 1.883 | 1.795 |
| 2500.00 | 1.700 | 1.612 |
| 1250.00 | 1.428 | 1.340 |
| 625.00 | 0.884 | 0.796 |
| 312.50 | 0.562 | 0.474 |
| 156.25 | 0.323 | 0.235 |
| 78.13 | 0.232 | 0.144 |
| 0.00 | 0.088 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 82-106% | 94% |
| EDTA plasma(n=5) | 92-127% | 110% |
| Heparin plasma(n=5) | 89-102% | 96% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 91-103% | 91-118% | 92-118% | 87-99% |
| EDTA plasma(n=5) | 83-100% | 92-104% | 84-108% | 89-119% |
| Heparin plasma(n=5) | 82-103% | 88-96% | 80-105% | 83-96% |