48 T | $320.00 |
96 T | $458.00 |
96 T*5 | $1947.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
10.00 | 1.960 | 1.867 |
5.00 | 1.654 | 1.561 |
2.50 | 1.214 | 1.121 |
1.25 | 0.941 | 0.848 |
0.63 | 0.513 | 0.420 |
0.31 | 0.320 | 0.227 |
0.16 | 0.191 | 0.098 |
0.00 | 0.093 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 82-111% | 97% |
EDTA plasma(n=5) | 88-124% | 106% |
Heparin plasma(n=5) | 83-89% | 86% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 86-107% | 91-99% | 93-99% | 87-98% |
EDTA plasma(n=5) | 82-103% | 82-113% | 83-115% | 80-107% |
Heparin plasma(n=5) | 84-103% | 85-115% | 94-119% | 82-94% |