| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 1.960 | 1.867 |
| 5.00 | 1.654 | 1.561 |
| 2.50 | 1.214 | 1.121 |
| 1.25 | 0.941 | 0.848 |
| 0.63 | 0.513 | 0.420 |
| 0.31 | 0.320 | 0.227 |
| 0.16 | 0.191 | 0.098 |
| 0.00 | 0.093 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 82-111% | 97% |
| EDTA plasma(n=5) | 88-124% | 106% |
| Heparin plasma(n=5) | 83-89% | 86% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-107% | 91-99% | 93-99% | 87-98% |
| EDTA plasma(n=5) | 82-103% | 82-113% | 83-115% | 80-107% |
| Heparin plasma(n=5) | 84-103% | 85-115% | 94-119% | 82-94% |