48 T | $320.00 |
96 T | $458.00 |
Concentration (pg/mL) | OD | Corrected OD |
---|---|---|
1000.00 | 2.204 | 2.106 |
500.00 | 1.657 | 1.559 |
250.00 | 1.088 | 0.990 |
125.00 | 0.782 | 0.684 |
62.50 | 0.539 | 0.441 |
31.25 | 0.397 | 0.299 |
15.63 | 0.206 | 0.108 |
0.00 | 0.098 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 88-102% | 95% |
EDTA plasma(n=5) | 79-97% | 88% |
Heparin plasma(n=5) | 83-95% | 89% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 89-98% | 81-94% | 96-105% | 92-103% |
EDTA plasma(n=5) | 88-97% | 86-97% | 85-103% | 89-97% |
Heparin plasma(n=5) | 79-96% | 85-99% | 79-93% | 89-97% |