Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
20.00 | 1.925 | 1.838 |
10.00 | 1.517 | 1.430 |
5.00 | 1.251 | 1.164 |
2.50 | 0.831 | 0.744 |
1.25 | 0.456 | 0.369 |
0.63 | 0.364 | 0.277 |
0.32 | 0.232 | 0.145 |
0.00 | 0.087 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 88-103% | 95% |
EDTA plasma(n=5) | 80-97% | 88% |
Heparin plasma(n=5) | 92-106% | 99% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 85-105% | 84-98% | 95-102% | 88-102% |
EDTA plasma(n=5) | 86-97% | 83-96% | 91-103% | 96-105% |
Heparin plasma(n=5) | 86-93% | 91-105% | 87-98% | 99-102% |