| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 500.00 | 2.179 | 2.084 |
| 250.00 | 1.782 | 1.687 |
| 125.00 | 1.205 | 1.110 |
| 62.50 | 0.864 | 0.769 |
| 31.25 | 0.539 | 0.444 |
| 15.63 | 0.324 | 0.229 |
| 7.82 | 0.196 | 0.101 |
| 0.00 | 0.095 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 92-106% | 99% |
| EDTA plasma(n=5) | 88-103% | 95% |
| Heparin plasma(n=5) | 77-95% | 86% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-95% | 85-98% | 82-97% | 92-101% |
| EDTA plasma(n=5) | 84-94% | 85-103% | 94-105% | 92-101% |
| Heparin plasma(n=5) | 81-94% | 96-105% | 92-103% | 82-96% |