| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 4000.00 | 1.946 | 1.859 |
| 2000.00 | 1.653 | 1.566 |
| 1000.00 | 1.132 | 1.045 |
| 500.00 | 0.958 | 0.871 |
| 250.00 | 0.561 | 0.474 |
| 125.00 | 0.436 | 0.349 |
| 62.50 | 0.211 | 0.124 |
| 0.00 | 0.087 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-99% | 95% |
| EDTA plasma(n=5) | 85-97% | 91% |
| Heparin plasma(n=5) | 85-97% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 82-103% | 81-91% | 85-101% | 81-90% |
| EDTA plasma(n=5) | 88-102% | 86-97% | 89-97% | 79-95% |
| Heparin plasma(n=5) | 93-102% | 81-96% | 93-106% | 82-95% |